9
presence of equipment which does not conform to
these specifications.
Place the PRO Monitor on a rigid, secure surface.
Monitor must only be used with mounting hardware,
poles, and stands recommended by GE Medical
Systems Information Technologies. See Appendix D.
The weight of the accessory basket contents should not
exceed 6.6 lb (3 kg).
Arrange the power cord, air hoses, and all cables
carefully so they do not constitute a hazard.
Verify calibration of BP parameter (temp and pulse
oximeter do not require calibration). Ensure that the
display is functioning properly before operating the
PRO Monitor.
Do not immerse the Monitor in water. If the Monitor is
splashed with water or becomes wet, wipe it
immediately with a dry cloth.
Do not gas sterilize or autoclave.
The PRO Monitor, when used with applied parts and
accessories approved by GE Medical Systems
Information Technologies, is protected against
defibrillator damage.
Notes
Waveforms may be distorted and readings inaccurate
when electrosurgical cautery equipment is used while
monitoring with the PRO Monitor.
The electromagnetic compatibility profile of the PRO
Monitor may change if accessories other than those
specified for use with the PRO Monitor are used.
Trend data are retained in the PRO Monitor when it is
turned off, except when the default is overridden by
selecting the Trend button under the Service menu.