Using the Menu System
Silence. Selection of this button will cause all alarms except
the FAILSAFE alarm to be muted. A confirmation menu will
appear in Area 2 of the LED. Selection of either Yes or No
returns the user to the Clinician mode menu. If silence is
confirmed, the Alarm Silence button (26) illuminates and
alarms are permanently muted. If silence is not confirmed,
the alarm will be audible.
Caution: Alarms will be muted until either the Monitor is
switched off and on again or the Alarm Silence button (26)
is pressed.
Main
. Selection of this button returns the user to the Main
menu.
Error and Warning Messages
The error panel appears in Area 2 of the LCD and indicates
the error and its code, if it has one. In this example, a limit
violation alarm (which has no error code) has occurred. A list
of alarm error messages and their codes is in Appendix B.
Alarm conditions are addressed in two ways: the Alarms
button and OK button.
Alarms Button
Selection of this button takes the user to the Alarms menu,
where the alarm limits can be adjusted. This button is
available only when a parameter alarm limit has been
violated.