82
Default Settings
SpO
2
(%) HIGH: 100
SpO
2
(%) LOW: 90
Sensitivity Mode 2 (for low perfusion-Default)
Averaging Time 12 seconds
FastSAT Mode 0 (for Off)
Masimo Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644;
5,769,785; 6,002,952; 6,036,642; 6,067,462; 6,206,830;
6,157,850, and international equivalents.
Mechanical
Dimensions Height: 9.8 in (25.0 cm)
Width: 9.8 in (24.8 cm)
Depth: 6.9 in (17.5 cm)
Weight, Including Battery 7.8 lb (3.5 kg)
Mountings Self-supporting on rubber feet
or pole mountable
Portability Carried by recessed handle or
pole mounted
Classification Information Mode of operation:
continuous
Degree of protection against
harmful ingress of water: Drip-
proof IPX1
Power Requirements
AC input voltage: 100-240
VAC, 50 / 60 Hz (nominal)
90 ~ 253 VAC, 47 ~ 63 Hz
(range), 50VA.
Protection against electrical
shock: Class 1
DC input voltage: 24 VDC
(nominal), 12-30 VDC, 36VA,
Commentaires sur ces manuels