use a disposable adhesive sensor that may tolerate more
motion.
Replace the sensor.
PROBLEM: The SpO
2
value or the pulse rate changes
rapidly; the pulse amplitude bar is erratic.
CAUSE:
Excessive patient motion may be making it impossible
for the SpO
2
function to find a pulse pattern.
An electrosurgical unit (ESU) may be interfering with
performance.
SOLUTION:
Check the patient.
If possible, keep the patient still; check whether the
sensor is applied securely and properly, and replace it if
necessary; move the sensor to a new site; use a sensor
that tolerates more motion.
If an ESU is interfering:
Move the SpO
2
cable as far from the ESU as possible.
Plug the Monitor and the ESU into different AC circuits.
Move the ESU ground pad as close to the surgical site as
possible.
The sensor may be damp or may need to be replaced
with a new sensor.
PROBLEM: The oxygen saturation measurement does not
correlate with the value calculated from a blood gas
determination.
CAUSE:
The SpO2 calculation may not have correctly adjusted
for the effects of pH; temperature; CO2; fetal
hemoglobin; or 2,3-DPG.
Accuracy can be affected by incorrect sensor application
or use; intravascular dyes; bright light; excessive patient
movement; venous pulsations; electrosurgical
interference; and placement of a sensor on an extremity
that has a blood pressure cuff, arterial catheter, or
intravascular line.