4-24 PRO 100-400 V2 Patient Monitor Service Manual Revision A
2009829-001
Calibration & Maintenance: SECTION 4. GENERAL MAINTENANCE
d. Enter the Service Mode access code 2 2 1 3.
e. Select NIBP.
8. Select Pop off.
9. The unit will display three zeros in the systolic and diastolic
windows.
10. Adjust VR2 (more to the middle of board) fully counter-clockwise.
11. Set the cuff pressure to 157 mmHg ±2 mmHg using your
connected NIBP test setup.
12. The air pressure within the system may fluctuate initially. Verify
pressure with your external manometer.
13. Adjust VR2 clockwise until the pressure is released and the
Monitor displays Systolic: 000 & Diastolic: 000.
14. Verify the overpressure setting by performing the Neonate
Overpressure Check 4.6.4.
15.Power the Monitor off.